7 Tips for Perfect pH

7 Tips for Perfect pH

There are 100’s of different applications for measuring pH. When you consider the number of available instruments, electrodes, calibration standards, and individual techniques, it might be hard to find two users that measure pH the exact same way. Some users are able to achieve the best possible results, while many more users struggle to get accurate and repeatable pH measurements.

Our 7 tips for perfect pH will help to take the guesswork out of your pH measurements, helping you get the most out of your pH system whether you have a pocket tester or an advanced benchtop meter.

Request your free 7 tips for perfect pH poster


Store your electrode properly in pH storage solution.

  • Never store in distilled, deionised, or tap water.
  • With refillable electrodes, close the filling hole for storage.
  • Keep the electrode upright if possible.

Item code Description
ORI810001W ROSS pH electrode storage solution, 475mL Login to buy
ORI910001W Orion pH electrode storage solution, 475 mL, 3M KCl buffered to pH 6.75 Login to buy
EUT01X211206W pH electrode storage solution 480mL, 4.0 M KCl at pH 6.82 Login to buy

Clean your electrode regularly

Cleaning a dirty, clogged or coated electrode can restore proper electrode performance and prolong the useful life of the electrode. Use the correct cleaning solution appropriate to the samples being measured.

Contaminant Cleaning instructions Related products
Organic oil/grease films Soak in liquid dish soap and warm water or diluted Neutracon for 5 min DECNEUTRACON-5L
Inorgaic salt deposits Soak probe in 0.1 M HCl for 2 min FSBJ/4350/15
Proteins Soak in a protein cleaning solution for 2 min EUT01X211216W
Biological contamination Soak in dilute sodium hypochlorite to disinfect for 2 min ROA3904/2.5L

Never re-use buffers

A pH system is only as accurate as the calibration buffers that are used for standardisation.

  • Calibrate regularly with fresh buffers.
  • pH 10 buffer in particular will drift in a matter of hours due to carbon dioxide absorption.
  • Check the expiry date on your buffers.

For best accuracy, use a minimum of two-point standardization; first with a buffer value close to the electrode systems zero potential (typically pH 7); and next with an acid or base buffer whose value brackets the expected pH value of the sample(s).

pH reference buffers by pH

pH 1
pH 2
pH 3
pH 4

pH 5
pH 6
pH 7
pH 8

pH 9
pH 10
pH 11
pH 12
pH 13

View all pH reference buffers ›

Rinse electrode tip well between readings

Use distilled water or electrode rinse solution to thoroughly rinse the electrode and never rub or blot the bulb to dry it. Gently shake off the excess water.

Electrode bulbs should always remain hydrated, and a dry electrode bulb will cause more problems than one or two drops of water in your sample.

Item code Description
BSPWL012.5 LabServ Pronalys Distilled Water, 5 L Login to buy
ORI911125W Electrode rinse solution pouches, 25pcs Login to buy
NAL2425-0505 Wash bottle, distilled water labelled, 500 mL, LPDE, Right-To-Know safety design Login to buy
NAL2428-0505 Wash bottle, distilled water labelled, 500 mL, LPDE, Right-To-Understand safety design Login to buy

Check electrode fill level

If you use a refillable electrode be sure to check your fill level before storing and measurement.

Ensure that the filling hole is closed for storage, but open for calibration and measurement.

Item code Description
ORI510011W Filling Solution for No Cal pH Electrodes, 60 mL, Pack of 5 Login to buy
ORI810007W Filling Solution for ROSS Ultra, ROSS, PerpHecT ROSS Electrodes, 60 mL, Pack of 5 Login to buy
ORI900001W Filling Solution for Single Junction Sure-Flow Reference Electrodes, 60 mL, Pack of 5 Login to buy
ORI900002W Double Junction Sure-Flow Reference Electrode, Inner Chamber, 60 mL, Pack of 5 Login to buy
ORI900003W Double Junction Sure-Flow Reference Electrode, Outer Chamber, 60 mL, Pack of 5 Login to buy
ORI900011W Standard, Speciality, PerpHecT Ag/AgCl, Single Junction Refillable Green and Micro pH Electrodes, ORP Electrodes, 60 mLPack of 5 Login to buy
ORI910008W Filling Solution for Double Junction, Double Junction KNIpHE™ Electrodes, 60 mL, Case of 5 Login to buy

Record temperature along with your pH

The temperature of your sample can have an effect on the pH. Ensure the temperature has stabilised on the probe before taking a reading and always record this along with the pH.

pH values of buffers at various temperatures
0˚C 1.67 4.00 6.98 7.11 9.46 10.32 12.79
5˚C 1.67 4.00 6.95 7.08 9.40 10.25 12.73
10˚C 1.67 4.00 6.92 7.06 9.33 10.18 12.67
20˚C 1.67 4.00 6.87 7.01 9.23 10.06 12.52
25˚C 1.68 4.01 6.86 7.00 9.18 10.01 12.46
30˚C 1.68 4.02 6.85 6.98 9.14 9.97 12.36
40˚C 1.69 4.04 6.84 6.97 9.07 9.89 12.17
50˚C 1.71 4.06 6.83 6.97 9.01 9.83 11.96
60˚C 1.72 4.09 6.84 6.97 8.96 9.79 11.73
70˚C 1.74 4.13 6.85 6.99 8.92 9.78 11.47
80˚C 1.77 4.16 6.86 7.03 8.89 9.78 11.19
90˚C 1.79 4.21 6.88 7.08 8.85 9.80 10.89

Check the pH slope and mV in pH 7 buffer regularly

The slope should be greater than 53.5 mV/pH or 92% and the mV for pH 7 buffer should be between -30 mV and +30 mV.

How to check your pH electrode slope

  1. Select the millivolt (mV) mode of your pH meter
  2. Using a pH electrode, obtain mV readings of two fresh calibration buffers (i.e., pH 4.0 and pH 7.0 are best)
  3. Determine the net mV change
  4. Determine the net mV change per pH unit change and compare using the chart below

Slope at 20°C at 25°C Status
1 pH unit 3 pH unit 1 pH unit 3 pH unit
>102 >59.3 >178.0 >60.3 >181.0 Poor
102% 59.3 178.0 60.3 181.0 OK
101% 58.8 176.3 59.8 179.3 Very Good
100% 58.2 174.5 59.2 177.5 Ideal
99% 57.6 172.8 58.6 175.7 Very Good
98% 57.0 171.0 58.0 173.9 Very Good
97% 56.4 169.3 57.4 172.2 Very Good
96% 55.8 167.5 56.8 170.4 Very Good
95% 55.3 165.8 56.2 168.6 Very Good
94% 57.0 171.0 58.0 173.9 OK
93% 54.1 162.3 55.0 165.1 OK
92% 53.5 160.5 54.4 163.3 OK
<92 <53.5 <160.5 <54.4 <163.3 Poor

Featured pH Products

Benchtop pH Meters

The ideal choice for dedicated pH testing in water, process, research, and QA/QC samples. The Thermo Scientific portfolio of pH Meters offer a range from simple & economical models to powerful & sophisticated pH meters, to help you analyze your samples with confidence.

Portable pH Meters

Our portable pH meters are designed to offer pH testing in the field for applications including environmental, drinking water, and wastewater analysis.

pH Electrodes

The Thermo Scientific electrodes portfolio spans a vast range of industries and applications, and are categorised into digestible classifications to help you choose which electrode is best for your application.